 TRUMPower
DESCRIPTION
100 WATT MEDICAL & ITE SWITCHING POWER SUPPLIES TMC100 SERIES
The TMC100 series of compact, open PCB constructed, AC-DC switching power supplies are capable of delivering 100 watts of continuous output power at convection. They operate at 90-264 VAC input voltage without the need of voltage selection, and are suited for medical, information technology and industrial applications. Approval to both EN 60601-1 and EN 60950-1 safety standards improves design-in time and reduces end equipment compliance costs.
FEATURES
Medical and ITE approvals Compact size 2" x 4" x 1.26" Ultra high power density 10 W/cubic inch Full range input 90-132 VAC, or 180-264 VAC Low earth leakage current Conducted EMI class B RoHS compliant
SAFETY STANDARD APPROVALS
INPUT SPECIFICATIONS
Input voltage: Input frequency: Input current: Earth leakage current: 90-132 VAC or 180-264 VAC 47-63 Hz 1.9A (rms) for 100-120 VAC 1.0A (rms) for 200-240 VAC 150 A max. @ 264 VAC, 63 Hz
UL 60601-1 CSA C22.2 No. 601.1
TUV EN60601-1
OUTPUT SPECIFICATIONS
Output voltage/current: Total output power: Ripple and noise: Overvoltage protection: Overcurrent protection: Temperature coefficient: Transient response: See rating chart 100 watts maximum 250 mV peak to peak on 5.0 V model , 1% peak to peak on other models. Provided on output; set at 110-140% of its nominal output voltage All outputs protected to short circuit conditions All outputs 0.04% / C maximum Maximum excursion of 4% or better on all models, recovering to 1% of final value within 500 s after a 25% step load change
o
GENERAL SPECIFICATIONS
Switching frequency: Efficiency: Hold-up time: Line regulation: Inrush current: Withstand voltage: 100 KHz 87-90% @ 230 VAC full load. 12 msec minimum at 110 VAC. 0.2% maximum at full load 40 A @ 115 VAC or 80 A @ 230 VAC, at 25 C cold start 4000 VAC from input to output 1500 VAC from input to ground 500 VAC from output to ground MTBF: EMC Performance EN55011 / EN55022: FCC: VCC: Class B conducted, class A radiated Class B conducted, class A radiated Class B conducted, class A radiated Harmonic distortion, class A Line flicker ESD, 8 KV air and 6 KV contact Radiated immunity, 3 V/m Fast transient/burst, 2 KV Surge, 1 KV diff., 2 KV com. Conducted immunity, 3 Vrms Magnetic field immunity, 3 A/m Voltage dips, 30% reduction for 500 ms 60% reduction for 100 ms >95% reduction for 5 sec. Performance criteria A, B, A. 270,000 hours at full load at 25 C ambient temperature, calculated per MIL-HDBK-217F
o o
ENVIRONMENTAL SPECIFICATIONS
Operating temperature: Storage temperature: Relative humidity: Derating: Cooling: -10 C to +70 C -40 C to +85 C 5% to 95% non-condensing Derate from 100% at +50 C linearly to 50% at +70 C Convection cooling
o o o o o o
EN61000-3-2: EN61000-3-3: EN61000-4-2: EN61000-4-3: EN61000-4-4: EN61000-4-5: EN61000-4-6: EN61000-4-8: EN61000-4-11:
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA Phone: 408-988-6616 Fax: 408-988-6622 email: sales@trumpower.com Website: www.TRUMPower.com
UNIVERSAL INPUT
OUTPUT VOLTAGE/CURRENT RATING CHART
P/N TMC100-S05 TMC100-S12 TMC100-S15 TMC100-S18 TMC100-S24 TMC100-S28 TMC100-S36 TMC100-S48 Vnom. 5V 12 V 15 V 18 V 24 V 28 V 36 V 48 V Output Imin. Imax. 0A 20 A 0A 8.34 A 0A 6.7 A 0A 5.56 A 0A 4.2 A 0A 3.58 A 0A 2.78 A 0A 2.1 A Tol. 2% 2% 2% 2% 2% 2% 2% 2%
TMC100 MEDICAL & ITE SERIES
Maximum Output Power 100 W 100 W 100 W 100 W 100 W 100 W 100 W 100 W
NOTES: 1. Safety approvals are for PCB form only. To order unit with cover fitted, add suffix "C".
MECHANICAL SPECIFICATIONS
NOTES: 1. 2. 3. Dimensions shown in inches [mm] Tolerance 0.02 [0.5] maximum Input connector mates with Molex housing 09-50-3031 and Molex 2878 series crimp terminal. Output connector mates with Molex housing 09-50-3061 and Molex 2878 series crimp terminal. To ensure compliance with level B emissions, connect the three " " marked mounting holes with metallic standoffs to chassis. Weight: 190 grams (0.44 lbs.) approx.
4.
5.
6. .
PIN CHART
PIN 1 MODEL TMC100-S05 TMC100-S18 TMC100-S36 TMC100-S12 TMC100-S24 TMC100-S48 TMC100-S15 TMC100-S28
RETURN RETURN RETURN OUTPUT OUTPUT OUTPUT
2
3
4
5
6
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA Phone: 408-988-6616 Fax: 408-988-6622 email: sales@trumpower.com Website: www.TRUMPower.com


▲Up To Search▲   

 